Please use this identifier to cite or link to this item: https://hdl.handle.net/11499/30331
Title: Assessment of abbott architect 25-oh vitamin D assay in different levels of vitamin D
Authors: Avcı, Esin
Demir, Süleyman
Aslan, Diler
Nar, Rukiye
Şenol, Hande
Keywords: 25-Hydroxyvitamin D
deficiency
insufficiency
method comparison
vitamin D
Publisher: Sciendo
Abstract: There is increasing requests of Vitamin D test in many clinical settings in recent years. However, immunoassay performance is still a controversial topic. Several diagnostic manufacturers have launched automated 25-hydroxyvitamin D (25-OH D) immunoassays in the past decade. We compared the performance of Abbott Architect 25-OH D Vitamin immunoassay with liquid chromatography-tandem mass spectrometry systems (LC-MS/MS) to evaluate immunoassay performance, especially in deficient groups. Eighty human serum samples were analyzed with Architect 25-OH D vitamin kit (Abbott Diagnostics, Lake Forest, IL, USA) and LC-MS/MS systems (Zivak Technology, Istanbul, Turkey). The results of the immunoassay method were compared with the LC-MS/MS using Passing-Bablok regression analysis, Bland-Altman plots and correlation coefficient analysis. We also evaluated results in four levels of D vitamin as a severe deficiency, deficiency, insufficiency, and sufficiency. Architect showed 9.59% bias from LC-MS/MS with smaller mean. Passing-Bablok regression analysis demonstrated the value of 0.95 slope and had a constant bias with an intercept value of -4.25. Concordance correlation coefficient showed moderate agreement with the value of 0.918 (95% CI 0.878-0.945). Two methods revealed good interrater agreement (kappa = 0.738). While the smallest bias determined in deficiency (9.95%) group, the biggest was in insufficiency (15.15%). Architect 25-OH D vitamin immunoassay can be used in routine measurements but had potential misclassification of vitamin D status in insufficient and deficient groups. Although there are recent standardization attempts in 25-OH D measurements, clinical laboratories must be aware of this method. © 2019 Esin Avci, Süleyman Demir, Diler Aslan, Rukiye Nar, Hande Şenol, published by Sciendo.
URI: https://hdl.handle.net/11499/30331
https://doi.org/10.2478/jomb-2019-0039
ISSN: 1452-8258
Appears in Collections:PubMed İndeksli Yayınlar Koleksiyonu / PubMed Indexed Publications Collection
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
Tıp Fakültesi Koleksiyonu
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Files in This Item:
File SizeFormat 
1452-82582001100A.pdf236.13 kBAdobe PDFView/Open
Show full item record



CORE Recommender

SCOPUSTM   
Citations

11
checked on Dec 28, 2024

WEB OF SCIENCETM
Citations

11
checked on Dec 20, 2024

Page view(s)

92
checked on Aug 24, 2024

Download(s)

40
checked on Aug 24, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.