Title: | Repeatability and reproducibility of Galilei measurements in normal keratoconic and postrefractive corneas | Authors: | Güler, E. Yağcı, Ramazan Akyol, M. Arslanyilmaz, Z. Balci, M. Hepşen, T.F. |
Keywords: | Galilei Keratoconus Repeatability Reproducibility adult analyzer anterior eccentricity anterior elevation best fit sphere anterior eye segment Article camera central corneal thickness controlled study cornea thickness female follow up human keratoconus major clinical study male measurement repeatability posterior corneal power posterior eccentricity posterior elevation best fit sphere priority journal prospective study refractive surgery reproducibility sex difference thinnest cornea thickness total corneal power visual system parameters cornea corneal pachymetry devices keratometry myopia pathology young adult Adult Cornea Corneal Pachymetry Corneal Topography Female Follow-Up Studies Humans Male Myopia Prospective Studies Refractive Surgical Procedures Reproducibility of Results Young Adult |
Publisher: | Elsevier | Abstract: | Objective: To assess the repeatability and reproducibility of the anterior segment measurements performed with a dual Scheimpflug analyzer (Galilei) in normal, keratoconic and post-refractive surgery corneas. Methods: To evaluate the repeatability, two additional measurements were performed by the first examiner. To assess reproducibility, this was later followed by a single reading by the second examiner. The following parameters were recorded and evaluated in this study; central corneal thickness (CCT), thinnest corneal thickness (TCT), mean total corneal power (TCP) in central (0-4mm), mean posterior corneal power (PCP) in central (0.5-2mm), anterior and posterior elevation (best fit sphere [BFS]) in central 8mm anterior and posterior eccentricity (?2) in central 8mm. Repeatability and reproducibility for each corneal parameter was assessed using the Bland-Altman analysis. Results: Each of the three groups was consisted of 20 subjects (totally 60 patients, 30 men and 30 women). The 95% LoA for repeatability was very small, indicating small discrepancies between measurements related to CCT. Acceptable repeatability was also achieved for the other parameters in each group. However, the 95% LoA for value TCP was larger in keratoconic eyes. The 95% LoA for reproducibility was also very small, and acceptable for all measured parameters in each group. In addition, the 95% LoA was larger for the measurement of CCT and TCT for postrefractive corneas. Conclusions: The anterior segment measurements provided by Galilei showed good repeatability and reproducibility for normal, keratoconic and postrefractive corneas. © 2014 British Contact Lens Association. | URI: | https://hdl.handle.net/11499/7832 https://doi.org/10.1016/j.clae.2014.04.004 |
ISSN: | 1367-0484 |
Appears in Collections: | PubMed İndeksli Yayınlar Koleksiyonu / PubMed Indexed Publications Collection Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection Tıp Fakültesi Koleksiyonu WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection |
Show full item record
CORE Recommender
SCOPUSTM
Citations
18
checked on Oct 13, 2024
WEB OF SCIENCETM
Citations
16
checked on Nov 14, 2024
Page view(s)
36
checked on Aug 24, 2024
Google ScholarTM
Check
Altmetric
Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.