Title: Repeatability and reproducibility of Galilei measurements in normal keratoconic and postrefractive corneas
Authors: Güler, E.
Yağcı, Ramazan
Akyol, M.
Arslanyilmaz, Z.
Balci, M.
Hepşen, T.F.
Keywords: Galilei
Keratoconus
Repeatability
Reproducibility
adult
analyzer
anterior eccentricity
anterior elevation best fit sphere
anterior eye segment
Article
camera
central corneal thickness
controlled study
cornea thickness
female
follow up
human
keratoconus
major clinical study
male
measurement repeatability
posterior corneal power
posterior eccentricity
posterior elevation best fit sphere
priority journal
prospective study
refractive surgery
reproducibility
sex difference
thinnest cornea thickness
total corneal power
visual system parameters
cornea
corneal pachymetry
devices
keratometry
myopia
pathology
young adult
Adult
Cornea
Corneal Pachymetry
Corneal Topography
Female
Follow-Up Studies
Humans
Male
Myopia
Prospective Studies
Refractive Surgical Procedures
Reproducibility of Results
Young Adult
Publisher: Elsevier
Abstract: Objective: To assess the repeatability and reproducibility of the anterior segment measurements performed with a dual Scheimpflug analyzer (Galilei) in normal, keratoconic and post-refractive surgery corneas. Methods: To evaluate the repeatability, two additional measurements were performed by the first examiner. To assess reproducibility, this was later followed by a single reading by the second examiner. The following parameters were recorded and evaluated in this study; central corneal thickness (CCT), thinnest corneal thickness (TCT), mean total corneal power (TCP) in central (0-4mm), mean posterior corneal power (PCP) in central (0.5-2mm), anterior and posterior elevation (best fit sphere [BFS]) in central 8mm anterior and posterior eccentricity (?2) in central 8mm. Repeatability and reproducibility for each corneal parameter was assessed using the Bland-Altman analysis. Results: Each of the three groups was consisted of 20 subjects (totally 60 patients, 30 men and 30 women). The 95% LoA for repeatability was very small, indicating small discrepancies between measurements related to CCT. Acceptable repeatability was also achieved for the other parameters in each group. However, the 95% LoA for value TCP was larger in keratoconic eyes. The 95% LoA for reproducibility was also very small, and acceptable for all measured parameters in each group. In addition, the 95% LoA was larger for the measurement of CCT and TCT for postrefractive corneas. Conclusions: The anterior segment measurements provided by Galilei showed good repeatability and reproducibility for normal, keratoconic and postrefractive corneas. © 2014 British Contact Lens Association.
URI: https://hdl.handle.net/11499/7832
https://doi.org/10.1016/j.clae.2014.04.004
ISSN: 1367-0484
Appears in Collections:PubMed İndeksli Yayınlar Koleksiyonu / PubMed Indexed Publications Collection
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
Tıp Fakültesi Koleksiyonu
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Show full item record



CORE Recommender

SCOPUSTM   
Citations

18
checked on Oct 13, 2024

WEB OF SCIENCETM
Citations

16
checked on Nov 14, 2024

Page view(s)

36
checked on Aug 24, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.