dc.description.abstractTo compare calculated and measured doses for different regions of anthropomorphic phantom and patients using ion chamber and thermoluminescence dosimetry (TLD) for total body irradiation. METHODS Measurements were done for lateral fields with 6 MV, gantry 82°, 40×40 cm2 field and 400 cm source-axis distance (SAD). Entrance-exit and midline doses were measured on anthropomorphic phantom by TLD and entrance-exit doses were measured by TLD and ion chamber on patients. RESULTS For anthropomorphic phantom measurements differences between calculated and measured entrance-exit doses of head, neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%, 4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%, -1.4% and 7.4% respectively. For patients; TLD differences were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%, -7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis measurements with ion chamber differences were within %0.1-1.9. CONCLUSION Total body irradiation is being applied in limit values in our clinic. © 2013 Association of Oncology.en_US
dc.language.isoenen_US
dc.relation.ispartofTurk Onkoloji Dergisien_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectIn-vivo dosimetryen_US
dc.subjectThermoluminescence dosimetryen_US
dc.subjectTotal body irradiationen_US
dc.subjectarticleen_US
dc.subjecthumanen_US
dc.subjectimaging phantomen_US
dc.subjectionization chamberen_US
dc.subjectpelvisen_US
dc.subjectradiation doseen_US
dc.subjectthermoluminescence dosimetryen_US
dc.subjectwhole body radiationen_US
dc.titleMidplane dose determination and verification of calculated doses in total body irradiationen_US
dc.typeArticleen_US
dc.identifier.volume28en_US
dc.identifier.issue2en_US
dc.identifier.startpage67
dc.identifier.startpage67en_US
dc.identifier.endpage74en_US
dc.identifier.doi10.5505/tjoncol.2013.898-
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.scopus2-s2.0-84882725060en_US
dc.identifier.wosWOS:000439266900003en_US
dc.identifier.scopusqualityQ4-
dc.ownerPamukkale University-
item.fulltextWith Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.cerifentitytypePublications-
item.languageiso639-1en-
item.openairetypeArticle-
item.grantfulltextopen-
Appears in Collections:Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
Tıp Fakültesi Koleksiyonu
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection
Files in This Item:
File SizeFormat 
pdf_TOD_858.pdf1.63 MBAdobe PDFView/Open
Show simple item record



CORE Recommender

SCOPUSTM   
Citations

2
checked on Nov 16, 2024

WEB OF SCIENCETM
Citations

2
checked on Nov 16, 2024

Page view(s)

26
checked on Aug 24, 2024

Download(s)

24
checked on Aug 24, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.