Title: Effect of analysis time on detection limit by energy dispersive X-ray fluorescence spectrometry
Authors: Ekinci, Raci
Ekinci, N.
Senemtasi, E.
Keywords: Detection limit
Energy dispersive X-ray fluorescence
Fluorescence intensity
Abstract: The influence of analysis time on detection limits in energy dispersive X-ray spectrometry was experimentally investigated. We utilized a high resolution Si(Li) detector (full width half maximum=160 eV at 5.9 keV) with a Camberra DSA-1000 desktop spectrum analyzer and Am-241 and Cd-109 radioactive point sources to collect X-ray spectra. The results showed that the detection limit decreased with increasing analysis time.