dc.description.abstract | Objectives: To assess the repeatability and reproducibility of anterior segment measurements including aberrometric measurements provided by a dual Scheimpflug analyzer (Galilei) system in normal eyes. Materials and Methods: Three repeated consecutive measurements were taken by two independent examiners. The following were evaluated: total corneal power and posterior corneal power, corneal higher-order wavefront aberrations (6.0 mm pupil), pachymetry at the central, paracentral, and peripheral zones, and anterior chamber depth (ACD). Repeatability was assessed by calculating the within-subject standard deviation, precision, repeatability, and intraclass correlation coefficient (ICC). Bland-Altman analysis was used for assessing reproducibility. Results: Thirty eyes of 30 patients were included. The best ICC values were for corneal pachymetry and ACD. For both observers, acceptable ICC was also achieved for the other parameters, the only exceptions being posterior corneal astigmatism and total high order aberration. The 95% LoA (Limits of Agreement) values for all measurements showed small variability between the two examiners. Conclusion: The Galilei system provided reliable measurements of anterior segment parameters. Therefore, the instrument can be confidently used for routine clinical use and research purposes. © 2015, Turkish Ophthalmology Society. All rights reserved. | en_US |
dc.language.iso | en | en_US |
dc.publisher | Turkish Ophthalmology Society | en_US |
dc.relation.ispartof | Turk Oftalmoloiji Dergisi | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | Galilei | en_US |
dc.subject | Interobserver repeatability | en_US |
dc.subject | Intraobserver | en_US |
dc.subject | Wavefront aberrations | en_US |
dc.subject | adult | en_US |
dc.subject | anterior eye chamber depth | en_US |
dc.subject | anterior eye segment | en_US |
dc.subject | anterior segment analysis system | en_US |
dc.subject | Article | en_US |
dc.subject | astigmatism | en_US |
dc.subject | controlled study | en_US |
dc.subject | cornea thickness | en_US |
dc.subject | corneal pachymetry | en_US |
dc.subject | corneal wavefront aberration | en_US |
dc.subject | correlation coefficient | en_US |
dc.subject | dual Scheimpflug analyzer | en_US |
dc.subject | female | en_US |
dc.subject | human | en_US |
dc.subject | keratometry | en_US |
dc.subject | male | en_US |
dc.subject | measurement repeatability | en_US |
dc.subject | prospective study | en_US |
dc.subject | refraction error | en_US |
dc.subject | reproducibility | en_US |
dc.subject | visual acuity chart | en_US |
dc.subject | anterior eye chamber | en_US |
dc.subject | human experiment | en_US |
dc.subject | measurement precision | en_US |
dc.subject | normal human | en_US |
dc.subject | observer variation | en_US |
dc.subject | posterior corneal astigmatism | en_US |
dc.subject | posterior corneal power | en_US |
dc.subject | total corneal astigmatism | en_US |
dc.subject | total corneal power | en_US |
dc.subject | visual system parameters | en_US |
dc.title | Repeatability and reproducibility of anterior segment measurements in normal eyes using dual scheimpflug analyzer | en_US |
dc.type | Article | en_US |
dc.identifier.volume | 45 | en_US |
dc.identifier.issue | 6 | en_US |
dc.identifier.startpage | 243 | |
dc.identifier.startpage | 243 | en_US |
dc.identifier.endpage | 248 | en_US |
dc.identifier.doi | 10.4274/tjo.16768 | - |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.identifier.pmid | 27800242 | en_US |
dc.identifier.scopus | 2-s2.0-84948734830 | en_US |
dc.identifier.wos | WOS:000219246900005 | en_US |
dc.identifier.scopusquality | Q4 | - |
dc.owner | Pamukkale University | - |
item.fulltext | With Fulltext | - |
item.openairecristype | http://purl.org/coar/resource_type/c_18cf | - |
item.cerifentitytype | Publications | - |
item.languageiso639-1 | en | - |
item.openairetype | Article | - |
item.grantfulltext | open | - |
crisitem.author.dept | 14.01. Surgical Medicine | - |
Appears in Collections: | PubMed İndeksli Yayınlar Koleksiyonu / PubMed Indexed Publications Collection Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection Tıp Fakültesi Koleksiyonu WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection |
Files in This Item:
File | Size | Format | |
---|---|---|---|
Ramazan Yağcı.pdf | 683.76 kB | Adobe PDF | View/Open |
CORE Recommender
SCOPUSTM
Citations
4
checked on Oct 13, 2024
WEB OF SCIENCETM
Citations
4
checked on Nov 21, 2024
Page view(s)
40
checked on Aug 24, 2024
Download(s)
10
checked on Aug 24, 2024
Google ScholarTM
Check
Altmetric
Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.