Please use this identifier to cite or link to this item: https://hdl.handle.net/11499/8022
Title: Midplane dose determination and verification of calculated doses in total body irradiation
Authors: Özdemir, Özlem
Hoca, S.
Olacak, N.
Özkök, S.
Kamer, S.
Anacak, Y.
Keywords: In-vivo dosimetry
Thermoluminescence dosimetry
Total body irradiation
article
human
imaging phantom
ionization chamber
pelvis
radiation dose
thermoluminescence dosimetry
whole body radiation
Abstract: To compare calculated and measured doses for different regions of anthropomorphic phantom and patients using ion chamber and thermoluminescence dosimetry (TLD) for total body irradiation. METHODS Measurements were done for lateral fields with 6 MV, gantry 82°, 40×40 cm2 field and 400 cm source-axis distance (SAD). Entrance-exit and midline doses were measured on anthropomorphic phantom by TLD and entrance-exit doses were measured by TLD and ion chamber on patients. RESULTS For anthropomorphic phantom measurements differences between calculated and measured entrance-exit doses of head, neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%, 4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%, -1.4% and 7.4% respectively. For patients; TLD differences were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%, -7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis measurements with ion chamber differences were within %0.1-1.9. CONCLUSION Total body irradiation is being applied in limit values in our clinic. © 2013 Association of Oncology.
URI: https://hdl.handle.net/11499/8022
https://doi.org/10.5505/tjoncol.2013.898
ISSN: 1300-7467
Appears in Collections:Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
Tıp Fakültesi Koleksiyonu
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Files in This Item:
File SizeFormat 
pdf_TOD_858.pdf1.63 MBAdobe PDFView/Open
Show full item record



CORE Recommender

SCOPUSTM   
Citations

2
checked on Sep 30, 2024

WEB OF SCIENCETM
Citations

2
checked on Sep 24, 2024

Page view(s)

26
checked on Aug 24, 2024

Download(s)

24
checked on Aug 24, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.